
Page 47 of 66
©Retrotec Inc. 2012
7.2 Serial Number and Calibration Date Check
It's recommended to check the serial number and calibration date on the DM-2 regularly.
If the screen does not display a valid serial number and calibration date, please contact Retrotec tech
support since your gauge may have switched over to default calibration values. This can be checked by
connected your gauge to another gauge, port to port to ensure they read within 1%.
7.3 Moving Tubes Cause Fluctuating Pressures
If the measured values on Channel A or B are fluctuating back and forth, in a way that doesn't
correspond with any of the current testing conditions, it's possible that the pressure tube is moving. A
moving tube can cause a significant error. For instance, a tube that is attached both to the gauge and
to the fan, but is in the path of the fan's air flow, can swing back and forth. This swinging motion can
easily cause fluctuations in the measured pressure.
It's never a good idea to leave a tube hanging in the air, either between two points, or out of a
window. The pressure tube should sit securely on the ground and be kept still as much as possible. The
effects of a moving tube are very similar to the effects of wind on a tube, which is described below.
To see the effect of a moving tube
1. Connect a pressure tube (10 feet or longer) to the blue port of the DM-2.
2. Set Time Averaging to one second.
3. Wave the end of the tube through the air. Notice the fluctuations.
4. Hold the end of the tube still, and swing the middle of the tube. Notice the fluctuations.
Figure 11. The effect of a moving tube
on pressure readings.
Komentarze do niniejszej Instrukcji