
Page 44 of 66
©Retrotec Inc. 2012
When performing a gauge calibration with two gauges, if one gauge is known to be calibrated
accurately, it can be used as a reference for the second gauge. In that case, the measured
pressure on the tested gauge should be within 1% of the measured pressure on the calibrated
gauge.
To perform a T-connection Fan Pressure check
1. Using a T-connection, connect a pressure tube to the Input port of
both Channel A and B.
2. Create a Fan Pressure on the end of the pressure tube.
3. Both Channel A and B should display the same measured pressure
(or within 1%).
A similar procedure can be performed, by using a syringe to create a Fan
Pressure. This can be done with either one or multiple gauges.
Although the procedure with a syringe is fairly widely used, it is not
recommended by Retrotec for the following reasons:
Requires equipment that most people don’t have (Magnehelic gauge).
Requires specific syringes (50 and 500 cc).
It's an awkward procedure.
Some part of the testing system of tubes and syringe will invariably be leaky and produce
inaccurate results.
To perform a syringe check
1. Connect two gauges and a syringe following the set up in Figure 5-1.
2. Connect a Magnehelic gauge as shown.
3. Slowly pull air out of the syringe. Pressure readings on both gauges should increase.
4. Both gauges should display the same measured pressure.
5. Repeat steps 1 - 3 on the Input and Reference ports of both gauges.
Figure 10. Syringe check setup
Figure 9. Fan Pressure check
using a T-connection check.
Komentarze do niniejszej Instrukcji