The Fan Speed Control output on the gauge cable uses RS-485 protocol. The Speed Control Cable can
thus extend approximately 1,200 metres (4,000 feet) between the gauge and the fan.
1.1.2 The DM-2 Screen
The Display Screen is where all measurement values are shown, as well as the current status of the
device, and test configuration.
Table 1: DM-2 Gauge Screen
Page for
more
information
Battery Health
/Baseline Value
Displays the battery health (degree of charge) for rechargeable batteries
only.
If “Baseline” has been activated, a pressure will appear that is deducted
from the “PrA” value.
Cycles between inactive as indicated by "- - - - " and "Hold" which will
hold the display and control functions until “Hold” is turned off. “Jog”
will be possible when using Set Speed or Set Pressure.
Displays the current value of the slope of the line along which
@Pressure estimates are extrapolated. Can be set between 0.50 and
1.00. A value of 0.65 is suggested for houses, 0.60 for ducts and 0.55 for
leaky buildings with large holes and for large buildings.
Displays the current Time Averaging setting: Off, 1s, 2s, 4s, 8s, 10s, 20s,
1m, 2m.
Displays the current Auto Zero status, On or Off.
Displays the current surface area, or enclosure volume setting, used for
calculations involving per unit area.
Displays the current fan speed as a percentage
Displays the current fan (also known as the Device). This must match
the fan that is in use.